Simultaneous Desalting and Selective Elution of Basic, Acidic, and Neutral Peptides
LC Conditions
Column
Synergi 4 µm Max-RP 80 Å, LC Column 150 x 4.6 mm, Ea
Brand
Synergi
Part No
00F-4337-E0
Phase Name
Max-RP
Elution Type
Gradient
Mobile Phase
A: 0.1% TFA in Water
B: 0.1% TFA in Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
90
10
2
12
35
65
Flow Rate
1.00 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C12 4 x 3.0mm, 10/Pk Part No.: AJ0-6074 - Holder Part No.: KJ0-4282
Sample Preparation Details
Description
Strata-X 33 µm Polymeric Reversed Phase, 100 mg / 6 mL, Tubes, 30/Pk
Part No
8B-S100-ECH
Condition 1
1.0mL Acetonitrile
Condition 2
1.0mL Water
Load
Load 0.10 mL Sample diluted in Water
Wash 1
1.0 mL Water
Elution 1
0.25mL Acetonitrile:Water
Elution 2
0.25mL Acetonitrile:Water
Elution 3
0.25mL Acetonitrile:1% Formic Acid
Evaporation
Evaporate to dryness
Reconstitute
0.60 mL Mobile phase
Order Items Used in This Application
- 1. HGRH Grag 14-26
- 2. Angiotensin I
- 3. Angiotensin II
- 4. Angiotensin III
- 5. Bradykinin
Similar Applications