Simultaneous Desalting and Selective Elution of Basic, Acidic, and Neutral Peptides

Simultaneous Desalting and Selective Elution of Basic, Acidic, and Neutral Peptides
LC Conditions
Column

Synergi 4 µm Max-RP 80 Å, LC Column 150 x 4.6 mm, Ea

Brand

Synergi

Part No

00F-4337-E0

Phase Name

Max-RP

Elution Type

Gradient

Mobile Phase

A: 0.1% TFA in Water

B: 0.1% TFA in Acetonitrile

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

90

10

2

12

35

65

Flow Rate

1.00 mL/min

Temperature

22°C

Detection

UV-Vis Abs.-Variable Wave.(UV) @ (22°C)

Detection Info
Sample Notes

SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C12 4 x 3.0mm, 10/Pk Part No.: AJ0-6074 - Holder Part No.: KJ0-4282

Sample Preparation Details
Description

Strata-X 33 µm Polymeric Reversed Phase, 100 mg / 6 mL, Tubes, 30/Pk

Part No

8B-S100-ECH

Condition 1

1.0mL Acetonitrile

Condition 2

1.0mL Water

Load

Load 0.10 mL Sample diluted in Water

Wash 1

1.0 mL Water

Elution 1

0.25mL Acetonitrile:Water

Elution 2

0.25mL Acetonitrile:Water

Elution 3

0.25mL Acetonitrile:1% Formic Acid

Evaporation

Evaporate to dryness

Reconstitute

0.60 mL Mobile phase

Order Items Used in This Application

Strata-X 33 µm Polymeric Reversed Phase, 100 mg / 6 mL, Tubes, 30/Pk
8B-S100-ECH

Strata-X 33 µm Polymeric Reversed Phase, 100 mg / 6 mL, Tubes, 30/Pk

View Product

Synergi 4 µm Max-RP 80 Å, LC Column 150 x 4.6 mm, Ea
00F-4337-E0

Synergi 4 µm Max-RP 80 Å, LC Column 150 x 4.6 mm, Ea

View Product

  • 1. HGRH Grag 14-26
  • 2. Angiotensin I
  • 3. Angiotensin II
  • 4. Angiotensin III
  • 5. Bradykinin
Similar Applications

No Similar applications found.