Patulin analysis using Gemini
LC Conditions
Column
Gemini 5 µm C18 110 Å, LC Column 150 x 3 mm, Ea
Brand
Gemini
Part No
00F-4435-Y0
Phase Name
C18
Elution Type
Isocratic
Mobile Phase
A: Water
B: Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
90
10
Flow Rate
0.50 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Gemini C18 4 x 2.0mm, 10/Pk Part No.: AJ0-7596 - Holder Part No.: KJ0-4282
Order Items Used in This Application
- 1. Patulin
-OEChem-06211003152D<br><br> 17 18 0 1 0 0 0 0 0999 V2000<br> 2.0000 -0.1550 0.0000 O 0 0 0 0 0 0 0 0 0 0 0 0<br> 4.6783 -1.4597 0.0000 O 0 0 0 0 0 0 0 0 0 0 0 0<br> 2.8660 1.3450 0.0000 O 0 0 0 0 0 0 0 0 0 0 0 0<br> 6.2619 -0.6550 0.0000 O 0 0 0 0 0 0 0 0 0 0 0 0<br> 3.7320 -0.1550 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0<br> 2.8660 0.3450 0.0000 C 0 0 3 0 0 0 0 0 0 0 0 0<br> 3.7320 -1.1550 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0<br> 2.0000 -1.1550 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0<br> 2.8660 -1.6550 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0<br> 4.6783 0.1497 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0<br> 5.2619 -0.6550 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0<br> 3.4030 0.6550 0.0000 H 0 0 0 0 0 0 0 0 0 0 0 0<br> 1.3894 -1.0473 0.0000 H 0 0 0 0 0 0 0 0 0 0 0 0<br> 1.7880 -1.7376 0.0000 H 0 0 0 0 0 0 0 0 0 0 0 0<br> 2.8660 -2.2750 0.0000 H 0 0 0 0 0 0 0 0 0 0 0 0<br> 4.8709 0.7391 0.0000 H 0 0 0 0 0 0 0 0 0 0 0 0<br> 2.3291 1.6550 0.0000 H 0 0 0 0 0 0 0 0 0 0 0 0<br> 1 6 1 0 0 0 0<br> 1 8 1 0 0 0 0<br> 2 7 1 0 0 0 0<br> 2 11 1 0 0 0 0<br> 3 6 1 0 0 0 0<br> 3 17 1 0 0 0 0<br> 4 11 2 0 0 0 0<br> 5 6 1 0 0 0 0<br> 5 7 1 0 0 0 0<br> 5 10 2 0 0 0 0<br> 6 12 1 0 0 0 0<br> 7 9 2 0 0 0 0<br> 8 9 1 0 0 0 0<br> 8 13 1 0 0 0 0<br> 8 14 1 0 0 0 0<br> 9 15 1 0 0 0 0<br> 10 11 1 0 0 0 0<br> 10 16 1 0 0 0 0<br>M END
Compound Name
Patulin
CID
4696
Molecular Formula
C7H6O4
Molecular Weight
154
No. Hydrogen Bond Acceptors
4
No. Hydrogen Bond Donors
1
Similar Applications
Reversed Phase
Mycotoxins in negative mode using Luna Omega Polar C18 50x2.1mm
Luna Omega 1.6µm Polar C18 100Å, LC Column 50 x 2.1 mm, EA
23574
Reversed Phase
Multi-Class Screening of 243 Mycotoxins by LC/MS/MS
Gemini 5 µm C18 110 Å, LC Column 150 x 4.6 mm, Ea
20027