Trichothecenes analysis using Gemini

Trichothecenes analysis using Gemini
LC Conditions
Column

Gemini 5 µm C18 110 Å, LC Column 150 x 3 mm, Ea

Brand

Gemini

Part No

00F-4435-Y0

Phase Name

C18

Elution Type

Gradient

Mobile Phase

A: 20mM Ammonium Acetate, pH 4.5

B: ACN

Gradient Profile

Step No.

Time(min)

%A

%B

1

0.5

90

10

2

15

20

80

Flow Rate

0.50 mL/min

Temperature

22°C

Detection

Mass Spectrometer (MS) @ (22°C)

Detection Info
Sample Notes

SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Gemini C18 4 x 2.0mm, 10/Pk Part No.: AJ0-7596 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Gemini 5 µm C18 110 Å, LC Column 150 x 3 mm, Ea
00F-4435-Y0

Gemini 5 µm C18 110 Å, LC Column 150 x 3 mm, Ea

View Product

  • 1. T-2
  • 2. Nivalenol
  • 3. Deoxynivalenol
  • 4. Fusarenon X
  • 5. 15-Acetyl-Deoxynivalenol
  • 6. 3-Acetyl-deoxynivalenol
  • 7. Diacetoxyscirpenol
  • 8. HT-2
Similar Applications

No Similar applications found.