Catecholamines using Synergi Fusion-RP
LC Conditions
Column
Synergi 4 µm Fusion-RP 80 Å, LC Column 150 x 4.6 mm, Ea
Brand
Synergi
Part No
00F-4424-E0
Phase Name
Fusion-RP
Elution Type
Gradient
Mobile Phase
A: 20mM Ammonium formate, pH 3.5
B: Methanol
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
97
3
2
10
65
35
Flow Rate
1.00 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Fusion-RP 4 x 3.0mm, 10/Pk Part No.: AJ0-7557 - Holder Part No.: KJ0-4282
Order Items Used in This Application
- 1. Norepinephrine
- 2. Epinephrine
- 3. Normetanephrine
- 4. Dopamine
- 5. Deoxyepinephrine
- 6. 3-Methoxytyramine
- 7. Serotonin
Similar Applications
No Similar applications found.