EPA Method 515.2: Chlorinated Acids on ZB-1701

EPA Method 515.2: Chlorinated Acids on ZB-1701
GC Conditions
Official Method

EPA 515

Column

Zebron ZB-1701, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea

Part No

7HG-G006-11

Phase Name

ZB-1701

Injection

Splitless @ 200°C

Carrier Gas

Constant Flow Helium, 1.2 mL/min

Detection

Electron Capture (ECD) @ (280°C)

Detection Info
Sample Notes

Analytes were methylated

Order Items Used in This Application

Zebron ZB-1701, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
7HG-G006-11

Zebron ZB-1701, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea

View Product

  • 1. 3,5-Dichlorobenzoic acid
  • 2. 2,4-Dichlorophenylacetic acid
  • 3. Dicamba
  • 4. Dichloroprop
  • 5. 4,4'-Dibromooctafluorobiphenyl
  • 6. 2,4-D
  • 7. Pentachlorophenol
  • 8. Silvex
  • 9. 2,4,5-T
  • 10. 2,4-DB
  • 11. Dinoseb
  • 12. Bentazon
  • 13. Dacthal
  • 14. Picloram
  • 15. Acifluorfen
Similar Applications