EPA Method 8100: Polynuclear Aromatic Hydrocarbons on ZB-5MS

EPA Method 8100: Polynuclear Aromatic Hydrocarbons on ZB-5MS
GC Conditions
Official Method

EPA 8100

Column

Zebron ZB-5MS, GC Cap. Column 10 m x 0.10 mm x 0.10 µm, Ea

Part No

7CB-G010-02

Phase Name

ZB-5MS

Injection

Split 20:1 @ 320°C

Carrier Gas

Constant Flow Helium, 0.6 mL/min

Detection

Flame Ionization (FID) @ (330°C)

Detection Info
Sample Notes

Analytes were 50ppm.

Order Items Used in This Application

Zebron ZB-5MS, GC Cap. Column 10 m x 0.10 mm x 0.10 µm, Ea
7CB-G010-02

Zebron ZB-5MS, GC Cap. Column 10 m x 0.10 mm x 0.10 µm, Ea

View Product

  • 1. Naphthalene
  • 2. 2-Methylnaphthalene
  • 3. Acenaphthalene
  • 4. Acenaphthene
  • 5. Fluorene
  • 6. Phenanthrene
  • 7. Anthracene
  • 8. Fluoranthene
  • 9. Pyrene
  • 10. Benz[a]anthracene
  • 11. Chrysene
  • 12. Benzo[b]fluoranthene
  • 13. Benzo[k]fluoranthene
  • 14. Benzo[a]pyrene
  • 15. Indeno[1,2,3-cd]pyrene
  • 16. Dibenz[a,h]anthracene
  • 17. Benzo[g,h,i]perylene
Similar Applications