MercuryMS - Luna 3µm C18(2) 20 x 2.0mm for Benzodiazepines in Plasma by LC/MS/MS
LC Conditions
Column
Luna 3 µm C18(2) 100 Å, LC Column 20 x 2 mm, MercuryMS, Ea
Brand
Luna
Part No
00M-4251-B0-CE
Phase Name
C18(2)
Elution Type
Gradient
Mobile Phase
A: 0.1% Formic Acid in Water
B: 0.1% Formic Acid in Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
85
15
2
1
5
95
3
1.01
85
15
4
2
85
15
Flow Rate
0.60 mL/min
Temperature
22°C
Detection
Mass Spectrometer (MS) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C8 4 x 2.0mm, 10/Pk Part No.: AJ0-4289 - Holder Part No.: KJ0-4282
Sample Preparation Details
Description
Strata-X 33 µm Polymeric Reversed Phase, 30 mg / 1 mL, Tubes, 100/Pk
Part No
8B-S100-TAK
Condition 1
1 mL Methanol at 1 mL/min
Condition 2
1 mL Water at 1 mL/min
Load
Load 1.00 mL Sample
Wash 1
1 mL Water at 1.5 mL/min
Wash 2
1 mL 20% MeOH in water at 1.5 mL/min
Elution 1
1 mL Methanol at 0.5 mL/min
Evaporation
Evaporate to dryness
Reconstitute
200 µL 15% CH3CN in water
Order Items Used in This Application
- 1. 7-Aminoflunitrazepam
- 2. Chlordiazepoxide
- 3. Oxazepam
- 4. Nordazepam
- 5. Clonazepam
- 6. Temazepam
- 7. Flunitrazepam
- 8. Diazepam
Similar Applications