MercuryMS - Luna 3µm C18(2) 20 x 2.0mm (mini column) for Tricyclics in Plasma by LC/MS/MS
LC Conditions
Column
Luna 3 µm C18(2) 100 Å, LC Column 20 x 2 mm, Ea
Brand
Luna
Part No
00M-4251-B0
Phase Name
C18(2)
Elution Type
Isocratic
Mobile Phase
A: 0.1% Formic Acid in Water
B: 0.1% Formic Acid in Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
28
72
Flow Rate
0.60 mL/min
Temperature
22°C
Detection
Mass Spectrometer (MS) @ (22°C)
Detection Info
Sample Notes
Total Ion Chromatogram is shown below. Single Ion Chromatograms are available. Please ask your Phenomenex Technical Representative.
Sample Preparation Details
Description
Strata-X-C 33 µm Polymeric Strong Cation, 30 mg / 1 mL, Tubes, 100/Pk
Part No
8B-S029-TAK
Condition 1
1 mL Methanol at 1 mL/min
Condition 2
1 mL Water at 1 mL/min
Load
Load 1.00 mL Sample
Wash 1
1 mL 0.1M acetic acid at 1.5 mL/min
Wash 2
2 mL MeOH at 1.5 mL/min
Elution 1
1 mL 5% ammonium hydroxide(28-30%) in MeOH at 0.5 mL/min
Evaporation
Evaporate to dryness
Reconstitute
200 µL Mobile Phase
Order Items Used in This Application
- 1. Doxepin
- 2. Desipramine
- 3. Imipramine
- 4. Nortriptyline
- 5. Trimipramine
- 6. Amitriptyline