Poly Brominated Diphenyl Ethers (PBDEs) on ZB-5HT 30x0.25x0.10

Poly Brominated Diphenyl Ethers (PBDEs) on ZB-5HT 30x0.25x0.10
GC Conditions
Column

Zebron ZB-5HT, GC Cap. Column 30 m x 0.25 mm x 0.10 µm, Ea

Part No

7HG-G015-02

Phase Name

ZB-5HT

Injection

Pulsed @ 350°C

Carrier Gas

Constant Flow Helium, 1.3 mL/min

Detection

Mass Selective (MSD) @ (0°C)

Detection Info
Sample Notes

Analytes were 2.5ppm in isooctane Pulsed splitless injection at 50psi using direct connect liner.

Order Items Used in This Application

Zebron ZB-5HT, GC Cap. Column 30 m x 0.25 mm x 0.10 µm, Ea
7HG-G015-02

Zebron ZB-5HT, GC Cap. Column 30 m x 0.25 mm x 0.10 µm, Ea

View Product

  • 1. BDE-25
  • 2. BDE-28
  • 3. BDE-75
  • 4. BDE-49
  • 5. BDE-71
  • 6. BDE-47
  • 7. BDE-66
  • 8. BDE-77
  • 9. BDE-100
  • 10. BDE-119
  • 11. BDE-99
  • 12. BDE-85
  • 13. BDE-154
  • 14. BDE-153
  • 15. BDE-138
  • 16. BDE-183
  • 17. BDE-190
  • 18. BDE-203
Similar Applications

No Similar applications found.