Simultaneous Desalting and Selective Elution of Basic, Neutral and Acidic Peptides
LC Conditions
Column
Synergi 4 µm Max-RP 80 Å, LC Column 150 x 4.6 mm, Ea
Brand
Synergi
Part No
00F-4337-E0
Phase Name
Max-RP
Elution Type
Gradient
Mobile Phase
A: 0.1% TFA in Water
B: 0.1% TFA in Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
90
10
2
12
35
65
Flow Rate
1.00 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C12 4 x 3.0mm, 10/Pk Part No.: AJ0-6074 - Holder Part No.: KJ0-4282
Sample Preparation Details
Description
Strata-X 33 µm Polymeric Reversed Phase, 30 mg / 1 mL, Tubes, 100/Pk
Part No
8B-S100-TAK
Condition 1
1 mL Acetonitrile
Condition 2
Water
Load
Load 840.00 µL Sample diluted in water with salt
Wash 1
2 mL Water
Elution 1
0.25 mL 40:60 Acetonitrile/Water. for acidic peptides
Elution 2
0.25 mL 70:30 Acetonitrile/Water. for neutral peptides
Elution 3
0.25 mL 70:30 Acetonitrile/1.0% Formic Acid in Water. for basic peptides
Evaporation
Evaporate to dryness
Reconstitute
20 µL initial mobile phase
Order Items Used in This Application
- 1. Adrenocorticotropin Hormone Fragment 1-24 (ACTH)
- 2. Bradykinin
- 3. Methionine enkephalin
- 4. Gonadotropin Releasing Hormone (HGRH) Fragment 14-26
Similar Applications