PEGylated Insulin on Jupiter® 300 C4, C18 and Two Experimental Columns
LC Conditions
Column
Jupiter 5 µm C4 300 Å, LC Column 150 x 4.6 mm, Ea
Brand
Jupiter
Part No
00F-4167-E0
Phase Name
C4
Elution Type
Gradient
Mobile Phase
A: 0.1% TFA and 2% ACN in Water
B: 70/20% ACN/IPA, 0.08% TFA in Water
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
85
15
2
25
30
70
Flow Rate
1.00 mL/min
Temperature
45°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
Reversed phase of PEGylated Insulin run on listed columns. Note the reduced resolution for the Jupiter C18 and EXP 2 columns when compared to the Jupiter C4 column. Although it is difficult to visualize, the Jupiter actully has better resolution of PEGylated forms than the EXP1 column.
Order Items Used in This Application
- 1. PEGylated Insulin
Similar Applications