PEGylated Insulin on Jupiter® 300 C4, C18 and Two Experimental Columns

PEGylated Insulin on Jupiter® 300 C4, C18 and Two Experimental Columns
LC Conditions
Column

Jupiter 5 µm C4 300 Å, LC Column 150 x 4.6 mm, Ea

Brand

Jupiter

Part No

00F-4167-E0

Phase Name

C4

Elution Type

Gradient

Mobile Phase

A: 0.1% TFA and 2% ACN in Water

B: 70/20% ACN/IPA, 0.08% TFA in Water

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

85

15

2

25

30

70

Flow Rate

1.00 mL/min

Temperature

45°C

Detection

UV-Vis Abs.-Variable Wave.(UV) @ (22°C)

Detection Info
Sample Notes

Reversed phase of PEGylated Insulin run on listed columns. Note the reduced resolution for the Jupiter C18 and EXP 2 columns when compared to the Jupiter C4 column. Although it is difficult to visualize, the Jupiter actully has better resolution of PEGylated forms than the EXP1 column.

Order Items Used in This Application

Jupiter 5 µm C4 300 Å, LC Column 150 x 4.6 mm, Ea
00F-4167-E0

Jupiter 5 µm C4 300 Å, LC Column 150 x 4.6 mm, Ea

View Product

  • 1. PEGylated Insulin
Similar Applications

No Similar applications found.