Jupiter® 300 C4 - Comparison of Tetrahydrofuran Organic Mixtures
LC Conditions
Column
Jupiter 5 µm C4 300 Å, LC Column 150 x 4.6 mm, Ea
Brand
Jupiter
Part No
00F-4167-E0
Phase Name
C4
Elution Type
Gradient
Mobile Phase
A: 0.1% TFA in Water
B: 90% ACN 0.09% TFA in Water
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
80
20
2
25
45
55
Flow Rate
1.00 mL/min
Temperature
45°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
Reversed phase runs of PEGylated beta lactoglobin using different organic mobile phase compositions on Jupiter 300 C4. 70% acetonitrile/ 20% tetrahydrofuran was substuted for the organic portion of the mobile phase. For various proteins tested the addition of THF slightly decreased the resolution of PEGylated forms.