Luna 5um C18(2) - HST Comparison

Luna 5um C18(2) - HST Comparison
LC Conditions
Column

Luna 5 µm C18(2) 100 Å, LC Column 250 x 4.6 mm, Ea

Brand

Luna

Part No

00G-4252-E0

Phase Name

C18(2)

Elution Type

Gradient

Mobile Phase

A: Water

B: Acetonitrile

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

90

10

2

1

90

10

3

25

5

95

4

35

5

95

5

35.1

90

10

Flow Rate

1.50 mL/min

Temperature

60°C

Detection

UV-Vis Abs.-Variable Wave.(UV) @ (22°C)

Detection Info
Sample Notes

SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Luna 5 µm C18(2) 100 Å, LC Column 250 x 4.6 mm, Ea
00G-4252-E0

Luna 5 µm C18(2) 100 Å, LC Column 250 x 4.6 mm, Ea

View Product

  • 1. Acetone
  • 2. 2-Butanone
  • 3. 2-Pentanone
  • 4. 2-Hexanone
  • 5. 2-Heptanone
  • 6. 2-Octanone
  • 7. 2-Decanone
  • 8. 2-Tridecanone
  • 9. 2-Hexadecanone
Similar Applications