Steroid Mix (11) on Luna 3um PFP(2)

Steroid Mix (11) on Luna 3um PFP(2)
LC Conditions
Column

Luna 3 µm PFP(2) 100 Å, LC Column 150 x 4.6 mm, Ea

Brand

Luna

Part No

00F-4447-E0

Phase Name

PFP(2)

Elution Type

Gradient

Mobile Phase

A: Water

B: Acetonitrile

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

70

30

2

25

35

65

Flow Rate

1.00 mL/min

Temperature

22°C

Detection

UV-Vis Abs.-Variable Wave.(UV) @ (22°C)

Detection Info
Sample Notes

SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, PFP 4 x 3.0 mm, 10/Pk Part No.: AJ0-8327 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Luna 3 µm PFP(2) 100 Å, LC Column 150 x 4.6 mm, Ea
00F-4447-E0

Luna 3 µm PFP(2) 100 Å, LC Column 150 x 4.6 mm, Ea

View Product

  • 1. Estriol
  • 2. Hydrocortisone
  • 3. Prednisone
  • 4. Cortisone
  • 5. Corticosterone
  • 6. Estradiol
  • 7. Cortisone acetate
  • 8. 21-Hydroxyprogesterone
  • 9. 17-Hydroxyprogesterone
  • 10. Estrone
  • 11. Progesterone
Similar Applications