Nicotine and Metabolites on Luna 5um HILIC

Nicotine and Metabolites on Luna 5um HILIC
LC Conditions
Column

Luna 5 µm HILIC 200 Å, LC Column 100 x 2 mm, Ea

Brand

Luna

Part No

00D-4450-B0

Phase Name

HILIC

Elution Type

Isocratic

Mobile Phase

A: 90/10 Acetonitrile/100mM Ammonium Formate pH 3.2

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

100

0

Flow Rate

0.40 mL/min

Temperature

22°C

Detection

Mass Spectrometer (MS) @ (22°C)

Detection Info
Sample Notes

The pH of the aqueous buffer is specified. The pH in mixed organic was not measured.

Order Items Used in This Application

Luna 5 µm HILIC 200 Å, LC Column 100 x 2 mm, Ea
00D-4450-B0

Luna 5 µm HILIC 200 Å, LC Column 100 x 2 mm, Ea

View Product

  • 1. Cotinine
  • 2. Nicotine
  • 3. Nornicotine
-OEChem-04281015372D<br><br> 25 26  0     1  0  0  0  0  0999 V2000<br>    2.6723    2.6739    0.0000 O   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.9511    0.9139    0.0000 N   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.6261   -2.1739    0.0000 N   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7601    0.3261    0.0000 C   0  0  1  0  0  0  0  0  0  0  0  0<br>    4.5691    0.9139    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.2601    1.8649    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.2601    1.8649    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7601   -0.6739    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.0000    0.6048    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.8940   -1.1739    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.6261   -1.1739    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.8940   -2.1739    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7601   -2.6739    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.3125    0.0446    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.8791    0.3769    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.1355    1.1660    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.8665    1.9938    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.1953    2.4815    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    1.8084    1.1945    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    1.4103    0.4132    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.1916    0.0152    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.3571   -0.8639    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.1630   -0.8639    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.3571   -2.4839    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7601   -3.2939    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>  1  7  2  0  0  0  0<br>  2  4  1  0  0  0  0<br>  2  7  1  0  0  0  0<br>  2  9  1  0  0  0  0<br>  3 11  1  0  0  0  0<br>  3 13  2  0  0  0  0<br>  4  5  1  0  0  0  0<br>  4  8  1  1  0  0  0<br>  4 14  1  0  0  0  0<br>  5  6  1  0  0  0  0<br>  5 15  1  0  0  0  0<br>  5 16  1  0  0  0  0<br>  6  7  1  0  0  0  0<br>  6 17  1  0  0  0  0<br>  6 18  1  0  0  0  0<br>  8 10  1  0  0  0  0<br>  8 11  2  0  0  0  0<br>  9 19  1  0  0  0  0<br>  9 20  1  0  0  0  0<br>  9 21  1  0  0  0  0<br> 10 12  2  0  0  0  0<br> 10 22  1  0  0  0  0<br> 11 23  1  0  0  0  0<br> 12 13  1  0  0  0  0<br> 12 24  1  0  0  0  0<br> 13 25  1  0  0  0  0<br>M  END
Compound Name

Cotinine

CID

854019

Molecular Formula

C10H12N2O

Molecular Weight

176

No. Hydrogen Bond Acceptors

2

Similar Applications

Reversed Phase

7.5 min Pain Panel on Kinetex 2.6µm Biphenyl, 30x2.1mm ESI+

Kinetex 2.6 µm Biphenyl 100 Å, LC Column 30 x 2.1 mm, Ea

23560

Reversed Phase

Fast Pain Panel ESI+

Kinetex 2.6 µm Biphenyl 100 Å, LC Column 30 x 2.1 mm, Ea

23384

Reversed Phase

Nicotine & Metabolites on Kinetex 2.6µm Evo C18 100x3mm

Kinetex 2.6 µm EVO C18 100 Å, LC Column 100 x 3.0 mm, EA

23006

Reversed Phase

Nicotines at High pH using Kinetex 5u EVO C18 100x3.0mm

Kinetex 5 µm EVO C18 100 Å, LC Column 100 x 3.0 mm, Ea

22721

Reversed Phase

Nicotine and metabolites by LCMS using Gemini NX 3u C18 100x2.1mm

Gemini 3 µm NX-C18 110 Å, LC Column 100 x 2 mm, Ea

20651

Reversed Phase

Nicotine and its Metabolites in Urine By LC/MS/MS Using Strata-X-C and Gemini-NX C18

Gemini 3 µm NX-C18 110 Å, LC Column 50 x 2 mm, Ea

19734

Reversed Phase

Nicotine and Metabolites on Gemini 5µm C18

Gemini 5 µm C18 110 Å, LC Column 100 x 2 mm, Ea

16345

Hydrophilic Interaction

Nicotine and Metabolites on Luna 5um HILIC

Luna 5 µm HILIC 200 Å, LC Column 100 x 2 mm, Ea

16335

Reversed Phase

Nicotine and Related Compounds

Synergi 4 µm Max-RP 80 Å, LC Column 150 x 4.6 mm, Ea

10901

Reversed Phase

Nicotine and Cotinine in Human Plasma

Luna 3 µm Phenyl-Hexyl 100 Å, LC Column 50 x 4.6 mm, Ea

1278