Hydrophobic Bases (Diltiazem, Promethazine) in Formic Acid on Gemini-NX 5um C18
LC Conditions
Column
Gemini 5 µm NX-C18 110 Å, LC Column 150 x 4.6 mm, Ea
Brand
Gemini
Part No
00F-4454-E0
Phase Name
NX-C18
Elution Type
Gradient
Mobile Phase
A: 0.1% formic acid in water
B: 0.1% formic acid in acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
95
5
2
10
5
95
Flow Rate
1.00 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282
Order Items Used in This Application
- 1. Diltiazem
- 2. Promethazine
Similar Applications