Hydrophobic Bases (Diltiazem, Promethazine) in Formic Acid on Gemini-NX 5um C18

Hydrophobic Bases (Diltiazem, Promethazine) in Formic Acid on Gemini-NX 5um C18
LC Conditions
Column

Gemini 5 µm NX-C18 110 Å, LC Column 150 x 4.6 mm, Ea

Brand

Gemini

Part No

00F-4454-E0

Phase Name

NX-C18

Elution Type

Gradient

Mobile Phase

A: 0.1% formic acid in water

B: 0.1% formic acid in acetonitrile

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

95

5

2

10

5

95

Flow Rate

1.00 mL/min

Temperature

22°C

Detection

UV-Vis Abs.-Variable Wave.(UV) @ (22°C)

Detection Info
Sample Notes

SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Gemini 5 µm NX-C18 110 Å, LC Column 150 x 4.6 mm, Ea
00F-4454-E0

Gemini 5 µm NX-C18 110 Å, LC Column 150 x 4.6 mm, Ea

View Product

  • 1. Diltiazem
  • 2. Promethazine
Similar Applications

No Similar applications found.