Tocainide on Lux 5µ Cellulose-1 250 x 4.6mm in NP

Tocainide on Lux 5µ Cellulose-1 250 x 4.6mm in NP
LC Conditions
Column

Lux 5 µm Cellulose-1, LC Column 250 x 4.6 mm, Ea

Brand

Lux

Part No

00G-4459-E0

Phase Name

Cellulose-1

Elution Type

Isocratic

Mobile Phase

A: 0.1 % DEA in Hexane

B: 0.1% DEA in Ethanol

Gradient Profile

Step No.

Time(min)

%A

%B

1

10

90

10

Flow Rate

1.00 mL/min

Temperature

22°C

Detection

UV-Vis Abs.-Variable Wave.(UV) @ (22°C)

Detection Info
Sample Notes

SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Lux Cellulose-1, 4 x 3.0mm, 10/Pk Part No.: AJ0-8403 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Lux 5 µm Cellulose-1, LC Column 250 x 4.6 mm, Ea
00G-4459-E0

Lux 5 µm Cellulose-1, LC Column 250 x 4.6 mm, Ea

View Product

  • 1. Tocainide
  • 2. Tocainide
-OEChem-05081000432D<br><br> 30 30  0     1  0  0  0  0  0999 V2000<br>    5.4641    0.1900    0.0000 O   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7320    0.1900    0.0000 N   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.4641    2.1900    0.0000 N   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7320   -0.8100    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.8660   -1.3100    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.5981   -1.3100    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.5981    1.6900    0.0000 C   0  0  3  0  0  0  0  0  0  0  0  0<br>    4.5981    0.6900    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.8660   -2.3100    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.5981   -2.3100    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.0000   -0.8100    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.4641   -0.8100    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7320   -2.8100    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7320    2.1900    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.5981    2.3100    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.1951    0.5000    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.3291   -2.6200    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.1350   -2.6200    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.3100   -0.2731    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    1.4631   -0.5000    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    1.6900   -1.3469    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.7741   -1.3469    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    6.0010   -0.5000    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.1541   -0.2731    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7320   -3.4300    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.0420    2.7269    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.1951    2.5000    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.4220    1.6531    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.4641    2.8100    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    6.0010    1.8800    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>  1  8  2  0  0  0  0<br>  2  4  1  0  0  0  0<br>  2  8  1  0  0  0  0<br>  2 16  1  0  0  0  0<br>  3  7  1  0  0  0  0<br>  3 29  1  0  0  0  0<br>  3 30  1  0  0  0  0<br>  4  5  2  0  0  0  0<br>  4  6  1  0  0  0  0<br>  5  9  1  0  0  0  0<br>  5 11  1  0  0  0  0<br>  6 10  2  0  0  0  0<br>  6 12  1  0  0  0  0<br>  7  8  1  0  0  0  0<br>  7 14  1  0  0  0  0<br>  7 15  1  0  0  0  0<br>  9 13  2  0  0  0  0<br>  9 17  1  0  0  0  0<br> 10 13  1  0  0  0  0<br> 10 18  1  0  0  0  0<br> 11 19  1  0  0  0  0<br> 11 20  1  0  0  0  0<br> 11 21  1  0  0  0  0<br> 12 22  1  0  0  0  0<br> 12 23  1  0  0  0  0<br> 12 24  1  0  0  0  0<br> 13 25  1  0  0  0  0<br> 14 26  1  0  0  0  0<br> 14 27  1  0  0  0  0<br> 14 28  1  0  0  0  0<br>M  END
Compound Name

Tocainide

CID

38945

Molecular Formula

C11H16N2O

Molecular Weight

192

No. Hydrogen Bond Acceptors

2

No. Hydrogen Bond Donors

2

Similar Applications

Tocainide on Lux 5µm Amylose-2 in NP

20090

Chiral

Tocainide on Lux 5µ Cellulose-1 250 x 4.6mm in NP

Lux 5 µm Cellulose-1, LC Column 250 x 4.6 mm, Ea

17676