Expanded Amphetamines by MSD on a ZB-Drug-1

Expanded Amphetamines by MSD on a ZB-Drug-1
GC Conditions
Column

Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea

Part No

7CD-G023-08

Phase Name

ZB-Drug-1

Injection

Split 10:1 @ 280°C

Carrier Gas

Constant Flow Helium, 55.0 cm/sec

Detection

Mass Selective (MSD) @ (230°C)

Detection Info
Sample Notes

Analytes are 40ppm derivatized with HFBA Sim ions: (Quant ion*) D11-Amphetamine (*244,128): Amphetamine (*240,118,117): D14-Methamphetamine (*261,213): Methamphetamine (*254,210,118): D5-MDA (*167,380,244): MDA (*162,375,240): D5-MDMA (*258,213,164): MDMA (*254,210,389): D5-MDEA (*273,241,408): MDEA (*268,240,403)

Order Items Used in This Application

Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
7CD-G023-08

Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea

View Product

  • 1. Amphetamine-d11
  • 2. Amphetamine
  • 3. Methamphetamine-d14
  • 4. Methamphetamine
  • 5. MDA-d5
  • 6. MDA
  • 7. MDMA-d5
  • 8. MDMA
  • 9. MDEA-d5
  • 10. MDEA
Similar Applications

No Similar applications found.