Amphetamines in Urine by LC/MS/MS on Gemini 5µm C18 150 x 3.0
LC Conditions
Column
Gemini 5 µm C18 110 Å, LC Column 150 x 3 mm, Ea
Brand
Gemini
Part No
00F-4435-Y0
Phase Name
C18
Elution Type
Isocratic
Mobile Phase
A: 10mM Ammonium formate w/0.1% formic acid
B: Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
90
10
Flow Rate
0.60 mL/min
Temperature
22°C
Detection
Mass Spectrometer (MS) @ (450°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Gemini C18 4 x 2.0mm, 10/Pk Part No.: AJ0-7596 - Holder Part No.: KJ0-4282
Sample Preparation Details
Description
Strata-X-C 33 µm Polymeric Strong Cation, 30 mg / 1 mL, Tubes, 100/Pk
Part No
8B-S029-TAK
Pretreatment Note
to prevent the loss of amphetamine and methamphetamine during solvent evaporation, add two drops of conc. HCl before the evaporation step
Condition 1
1 mL Methanol
Condition 2
1 mL Water
Load
Load Sample
Wash 1
1 mL Water
Wash 2
1 mL 0.1M Acetic Acid
Wash 3
1 mL Methanol
Elution 1
1 mL 2% of ammonium hydroxide (28-30%) in methanol
Evaporation
Evaporate to dryness
Reconstitute
200 µL Mobile Phase
Order Items Used in This Application
- 1. 4-Hydroxy-3-methoxymethamphetmine
- 2. Amphetamine
- 3. Methamphetamine
- 4. 3,4-Methylenedioxyamphetamine (MDA)-d5
- 5. 3,4-Methylenedioxyamphetamine (MDA)
- 6. 3,4-Methylendioxymethamphetamine-d5 (MDMA)-d5
- 7. 3,4-Methylendioxmethamphetamine (MDMA)
Similar Applications