THC and Metabolite in Plasma by LC/MS/MS on Gemini C18 5µm 50 x 3.0
LC Conditions
Column
Gemini 3 µm C18 110 Å, LC Column 50 x 3 mm, Ea
Brand
Gemini
Part No
00B-4439-Y0
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 10mM NH4HCO3 (pH 10)
B: Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0.1
85
15
2
3
5
95
3
3.5
5
95
4
3.6
85
15
5
4.1
85
15
Flow Rate
0.80 mL/min
Temperature
22°C
Detection
Mass Spectrometer (MS) @ (450°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Gemini C18 4 x 2.0mm, 10/Pk Part No.: AJ0-7596 - Holder Part No.: KJ0-4282
Sample Preparation Details
Description
Strata-X-C 33 µm Polymeric Strong Cation, 30 mg / 1 mL, Tubes, 100/Pk
Part No
8B-S029-TAK
Condition 1
1 mL Methanol
Condition 2
1 mL DI Water
Load
Load 2.00 mL Sample diluted in 1mL Plasma diluted 1:1 with Water
Wash 1
1 mL 0.1N HAC
Wash 2
1 mL 30% ACN in 0.1N HAC
Wash 3
0.2
Elution 1
1 mL 2% HAC in CH3CN by gravity
Evaporation
Evaporate to dryness
Reconstitute
100 µL 15:85/ACN:10mM NH4HCO3, pH=10.0
Order Items Used in This Application
- 1. 11-nor-9-Carboxy-THC (THC-COOH)
- 2. delta-9-tetrahydrocannabinol-d3 (THC)
- 3. delta-9-tetrahydrocannabinol (THC)
Similar Applications