Organochloropesticides (OCP) and Polychlorinated Biphenyls (PCB) in Milk and Milk Products by GC/ECD

Organochloropesticides (OCP) and Polychlorinated Biphenyls (PCB) in Milk and Milk Products by GC/ECD
GC Conditions
Column

Zebron ZB-MultiResidue-1, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea

Part No

7HG-G016-11

Phase Name

ZB-MultiResidue-1

Injection

Splitless @ 50°C

Carrier Gas

Constant Flow Hydrogen, 1.5 mL/min

Detection

Electron Capture (ECD) @ (300°C)

Detection Info
Sample Notes

The application was kindly provided by the MLF, Dr.-Oskar-Farny-Institute in Wangen, Germany. Injector Gerstel-KAS3. Program: 50 °C with 12 °C/s to 300 °C, 3 min hold Solvent Venting Purge Time: 0.15 min, Splitless Time: 1 min, Equilibration Time 0.5 min, Initial Time: 0.15 min. Concentration: 25 ppb each for OCP + PCB, 100 ppb for Nitrofen

Order Items Used in This Application

Zebron ZB-MultiResidue-1, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
7HG-G016-11

Zebron ZB-MultiResidue-1, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea

View Product

  • 1. alpha-HCH
  • 2. Hexachlorobenzene
  • 3. gamma-HCH
  • 4. beta-HCH
  • 5. delta-HCH
  • 6. PCB 28
  • 7. PCB 52
  • 8. Aldrin
  • 9. Heptachloro epoxide
  • 10. 2,4'-DDE (o,p-)
  • 11. PCB 101
  • 12. 4,4'-DDE (p,p-)
  • 13. Nitrofen
  • 14. Dieldrin
  • 15. 2,4'-DDD (o,p-)
  • 16. Endrin
  • 17. 4,4'-DDD (p,p-)
  • 18. PCB 153
  • 19. PCB 138
  • 20. PCB 180
  • 21. Mirex
Similar Applications

No Similar applications found.