Interferon-Alpha (Intact & Unfolded) on Jupiter 3u C18 and Jupiter 5u C4
LC Conditions
Column
Jupiter 3 µm C18 300 Å, LC Column 150 x 2 mm, Ea
Brand
Jupiter
Part No
00F-4263-B0
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 0.1% TFA and 2% Acetonitrile in Water
B: 0.085% TFA, 90% Acetonitrile in Water
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
80
20
2
10
20
80
3
15
10
90
Flow Rate
0.30 mL/min
Temperature
25°C
Detection
UV-Vis Abs.-Diode Array (PDA) @ (25°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Widepore C18 4 x 2.0mm, 10/Pk Part No.: AJ0-4320 - Holder Part No.: KJ0-4282
Order Items Used in This Application
- 1. alpha-Interferon
Similar Applications