Steroids in Standard Urine on Gemini 5µm C18 150 x 3mm ID
LC Conditions
Column
Gemini 5 µm C18 110 Å, LC Column 150 x 3 mm, Ea
Brand
Gemini
Part No
00F-4435-Y0
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 0.1 % Formic Acid in Water
B: 0.1 % Formic Acid in 95% Acetonitrile/Water
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
60
40
2
4
100
0
3
15
25
75
4
15.1
40
60
5
20
40
60
Flow Rate
0.60 mL/min
Temperature
22°C
Detection
Mass Spectrometer (MS) @ (450°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Gemini C18 4 x 2.0mm, 10/Pk Part No.: AJ0-7596 - Holder Part No.: KJ0-4282
Sample Preparation Details
Description
Strata-X 33 µm Polymeric Reversed Phase, 30 mg / 1 mL, Tubes, 100/Pk
Part No
8B-S100-TAK
Condition 1
1 mL Methanol
Condition 2
1 mL Water
Load
Load 1.00 mL Sample diluted in Human serum or urine
Wash 1
1 mL Water
Wash 2
1 mL 40% Methanol in Water
Elution 1
1 mL Methanol
Evaporation
Evaporate to dryness
Reconstitute
100 µL 100 uL acetonitrile/water 40:60
Order Items Used in This Application
- 1. Triamcinolone
- 2. Prednisolone
- 3. Cortisone
- 4. Betamethasone
- 5. Corticosterone
- 6. Triamcinolone acetonide
- 7. 11-alpha-Hydroxyprogesterone
- 8. Cortisone acetate
- 9. 11-Ketoprogesterone
- 10. Betamethasone 17-valerate
Similar Applications
No Similar applications found.