Propoxyphene from urine by Screen C on a ZB-Drug1
GC Conditions
Column
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Part No
7CD-G023-08
Phase Name
ZB-Drug-1
Injection
Split 10:1 @ 260°C
Carrier Gas
Constant Flow Helium, 55.0 cm/sec
Detection
Mass Selective (MSD) @ (320°C)
Detection Info
Sample Notes
Analytes are 300ng/mL in urine Sim ions: (Quant ion*) D11-Propoxyphene (*64,198,213): Propoxyphene (*58,193,208)
Sample Preparation Details
Condition 1
3 mL Methanol at 1 mL/min
Condition 2
3 mL Water at 1 mL/min
Condition 3
1 mL 100mM phosphate buffer (pH 6.0) at 1 mL/min
Sample Pretreatment
6mL/200mg
Load
Load 2.00 mL Sample diluted in Urine
Wash 1
3 mL Water at 1 mL/min
Wash 2
1 mL 100mM Acetic acid at 1 mL/min
Wash 3
3 mL Methanol at 1 mL/min
Dry
5 minutes at > 10 inch Hg
Elution 1
3 mL MeCl2NH3OH (78:20:2) at 1 mL/min
Evaporation
Evaporate to dryness
Reconstitute
50 µL Ethyl acetate
Order Items Used in This Application
- 1. Propoxyphene-d11
- 2. Propoxyphene
Similar Applications
No Similar applications found.