Methadone from urine by Screen C on a ZB-Drug1

Methadone from urine by Screen C on a ZB-Drug1
GC Conditions
Column

Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea

Part No

7CD-G023-08

Phase Name

ZB-Drug-1

Injection

Split 20:1 @ 280°C

Carrier Gas

Constant Flow Helium, 55.0 cm/sec

Detection

Mass Selective (MSD) @ (320°C)

Detection Info
Sample Notes

Analytes are 300ng/mL in urine Sim ions: (Quant ion*) D9-Methadone (*78,303): Methadone (*72,294,223)

Sample Preparation Details
Condition 1

3 mL Methanol at 1 mL/min

Condition 2

3 mL Water at 1 mL/min

Condition 3

2 mL 100mM Phosphate buffer (pH6.0) at 1 mL/min

Load

Load 2.00 mL Sample diluted in Urine

Wash 1

3 mL Water at 1 mL/min

Wash 2

1 mL 100mM Acetic acid at 1 mL/min

Wash 3

3 mL Methanol at 1 mL/min

Dry

5 minutes at > 10 inch Hg

Elution 1

MeCl2NH3OH (78:20:2)

Evaporation

Evaporate to dryness

Reconstitute

50 µL Ethyl acetate

Order Items Used in This Application

Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
7CD-G023-08

Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea

View Product

  • 1. Methadone-d9
  • 2. Methadone