6-MAM from urine by Screen C on a ZB-Drug1
GC Conditions
Column
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Part No
7CD-G023-08
Phase Name
ZB-Drug-1
Injection
Split 10:1 @ 260°C
Carrier Gas
Constant Flow Helium, 0.7 mL/min
Detection
Mass Selective (MSD) @ (300°C)
Detection Info
Sample Notes
Analytes are 10ng/mL in urine derivatized with BSTFA Sim ions: (Quant ion*) D3-6-MAM (*402,403): 6-MAM (*399,400,340)
Sample Preparation Details
Condition 1
3 mL Methanol at 1 mL/min
Condition 2
3 mL Water at 1 mL/min
Condition 3
1 mL 100mM phosphate buffer (pH6.0) at 1 mL/min
Load
Load 4.00 mL Sample diluted in Urine
Wash 1
3 mL Water at 1 mL/min
Wash 2
2 mL 100mM acetate buffer (pH 4.5) at 1 mL/min
Wash 3
3 mL Methanol at 1 mL/min
Dry
10minutes at > 10 inch Hg
Elution 1
3 mL MeCl2NH3OH (78:20:2) at 1 mL/min
Evaporation
Evaporate to dryness
Reconstitute
50 µL BSTFA
Order Items Used in This Application
- 1. 6-MAM-d3
- 2. 6-MAM
Similar Applications
No Similar applications found.