Expanded amphetamines and interferences including TFMPP by MSD on a ZB-Drug1
GC Conditions
Column
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Part No
7CD-G023-08
Phase Name
ZB-Drug-1
Injection
Split 10:1 @ 260°C
Carrier Gas
Constant Flow Helium, 0.5 mL/min
Detection
Mass Selective (MSD) @ (320°C)
Detection Info
Sample Notes
Analytes are 20ppm derivatized with HFBA Sim ions: (Quant ion*) D11-Amphetamine (*244,128): Amphetamine (*240,118,117): D14-Methamphetamine (*261,213): Methamphetamine (*254,210,118): D5-MDA (*167,380,244): MDA (*162,375,240): D5-MDMA (*258,213,164): MDMA (*254,210,389): D5-MDEA (*273,241,408): MDEA (*268,240,403) Note: Amphetamine-d11 does not share ions with coeluting interferences.
Order Items Used in This Application
- 1. Norephedrine
- 2. Norpseudoephedrine
- 3. Phentermine
- 4. Amphetamine-d11
- 5. Amphetamine
- 6. Ephedrine
- 7. Methamphetamine-d14
- 8. Methamphetamine
- 9. Pseudoephedrine
- 10. MDA-d5
- 11. MDA
- 12. MDMA-d5
- 13. MDMA
- 14. TFMPP
- 15. MDEA-d5
- 16. MDEA
Similar Applications