Benzodiazepines from urine by Screen C on a ZB-Drug1
GC Conditions
Column
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Part No
7CD-G023-08
Phase Name
ZB-Drug-1
Injection
Split 10:1 @ 260°C
Carrier Gas
Constant Flow Helium, 0.7 mL/min
Detection
Mass Selective (MSD) @ (320°C)
Detection Info
Sample Notes
Analytes are 300ng/mL in urine derivatized with MTBSTFA Sim ions: (Quant ion*) D5-Nordiazepam (*332,334): Nordiazepam (*327,328,383): D5-Oxazepam (*462,519): Oxazepam (*457,513,459): D5-Temazepam (*362,288,390): Temazepam (*357,359,385): Lorazepam (*491,513,493): Clonazepam (*372,374,326):Diazepam (*256,283,221): Alprazolam (*279,308,204): D5-a-Hydroxyalprazolam (*386,388): a-Hydroxyalprazolam (*381,383,384)
Sample Preparation Details
Condition 1
3 mL Methanol at 1 mL/min
Condition 2
3 mL Water at 1 mL/min
Condition 3
1 mL 100mM phosphate buffer (pH 6.0) at 1 mL/min
Load
Load 2.00 mL Sample diluted in Urine
Wash 1
2 mL Water at 1 mL/min
Wash 2
2 mL 20% acetonitrile in 100mM phosphate buffer (pH 6.0) at 1 mL/min
Dry
Dry column(5 minutes at > 10 in Hg) add 2mL of hexane and aspirate
Elution 1
5 mL Ethyl acetate at 1 mL/min
Evaporation
Evaporate to dryness
Reconstitute
50 µL Pyridine/MTBSTFA
Order Items Used in This Application
- 1. Nordiazepam-d5
- 2. Nordiazepam
- 3. Diazepam
- 4. Oxazepam-d5
- 5. Oxazepam
- 6. Lorazepam
- 7. Temazepam-d5
- 8. Temazepam
- 9. Clonazepam
- 10. Alprazolam
- 11. alpha-Hydroxyalprazolam-d5
- 12. alpha-Hydroxyalprazolam
Similar Applications