Ion Pairing & LC/MS Influence of Oligonucleotides (Figure 6c)
LC Conditions
Column
Clarity 3 µm Oligo-RP, LC Column 100 x 2 mm, Ea
Brand
Clarity LC
Part No
00D-4441-B0
Phase Name
Oligo-RP
Elution Type
Gradient
Mobile Phase
A: 200mM HFIP/8mMTEA/2% MeOH
B: 200mM HFIP/8mMTEA/98% MeOH
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
95
5
2
5
90
10
3
10
85
15
4
60
70
30
Flow Rate
0.25 mL/min
Temperature
50°C
Detection
Mass Spectrometer (MS) @ (365°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Clarity Oligo-RP 4 x 2.0mm, 10/Pk Part No.: AJ0-8134 - Holder Part No.: KJ0-4282
Order Items Used in This Application
- 1. poly dT (12-18) ladder
Similar Applications