Poly dT 12-18 standard on Oligo-RP (modifying TEA/HFIP Ratio)
LC Conditions
Column
Clarity 3 µm Oligo-RP, LC Column 150 x 2 mm, Ea
Brand
Clarity LC
Part No
00F-4441-B0
Phase Name
Oligo-RP
Elution Type
Gradient
Mobile Phase
A: TEAA/HFIP/2%MeOH in Water
B: Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
100
0
2
10
90
10
3
30
70
30
Flow Rate
0.20 mL/min
Temperature
22°C
Detection
Mass Spectrometer (MS) @ (22°C)
Detection Info
Sample Notes
Different ratios of HFIP to TEA were investigated using the poly dT 12-18 standard mixture. HFIP was fixed at 200 mM and different levels of TEA were used between 1 and 20 mM (4mm-red trace {pH 7.6}, 8mm-blue trace{pH8.0} and 14mm -black trace{pH8.3}). Optimal retention, MS response, and resolution were observed for the 200 mM HFIP./ 8mM TEA mobile phase mixture.