EPA Method 8330: Explosive Mix Using Synergi Hydro-RP

EPA Method 8330: Explosive Mix Using Synergi Hydro-RP
LC Conditions
Column

Synergi 4 µm Hydro-RP 80 Å, LC Column 250 x 4.6 mm, Ea

Brand

Synergi

Part No

00G-4375-E0

Phase Name

Hydro-RP

Elution Type

Gradient

Mobile Phase

A: Methanol

B: 20mM KH2PO4-pH=2.5

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

55

45

2

2

55

45

3

12

65

35

4

22

65

35

Flow Rate

1.00 mL/min

Temperature

25°C

Detection

UV-Vis Abs.-Variable Wave.(UV) @ (22°C)

Detection Info
Sample Notes

100ul of 100ug/ml 17 component mix was diluted with 900ul of 50:50 Methanol/water

Order Items Used in This Application

Synergi 4 µm Hydro-RP 80 Å, LC Column 250 x 4.6 mm, Ea
00G-4375-E0

Synergi 4 µm Hydro-RP 80 Å, LC Column 250 x 4.6 mm, Ea

View Product

  • 1. Octahydro-1,3,5,7-tetranitro-1,3,5,7-tetrazocine (HMX)
  • 2. Hexahydro-1,3,5-trinitro-1,3,5-triazine
  • 3. Picric acid
  • 4. 1,3,5-Trinitrobenzene (1,3,5-TNB)
  • 5. Methyl-2,4,6-trinitrophenylnitramine
  • 6. 1,3-Dinitrobenzene (1,3-DNB)
  • 7. Nitrobenzene
  • 8. Nitroglycerin
  • 9. 2,4,6-Trinitrotoluene (2,4,6-TNT)
  • 10. 3,5-Dinitroaniline
  • 11. 4-Amino-2,6-dinitrotoluene
  • 12. 2-Amino-4,6-dinitrotoluene
  • 13. 2,6-Dinitrotoluene (2,6-DNT)
  • 14. 2,4-Dinitrotoluene (2,4-DNT)
  • 15. Pentaerythritol tetranitrate (PETN)
  • 16. 2-Nitrotoluene
  • 17. 4-Nitrotoluene
  • 18. 3-Nitrotoluene