EPA Method 8330: Explosive Mix Using Luna Phenyl Hexyl

EPA Method 8330: Explosive Mix Using Luna Phenyl Hexyl
LC Conditions
Column

Luna 5 µm Phenyl-Hexyl 100 Å, LC Column 250 x 4.6 mm, Ea

Brand

Luna

Part No

00G-4257-E0

Phase Name

Phenyl-Hexyl

Elution Type

Gradient

Mobile Phase

A: Methanol

B: Water

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

55

45

2

2

55

45

3

22

75

25

4

22.01

55

45

5

25

55

45

Flow Rate

1.00 mL/min

Temperature

22°C

Detection

UV-Vis Abs.-Variable Wave.(UV) @ (22°C)

Detection Info
Sample Notes

SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Phenyl 4 x 3.0mm, 10/Pk Part No.: AJ0-4351 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Luna 5 µm Phenyl-Hexyl 100 Å, LC Column 250 x 4.6 mm, Ea
00G-4257-E0

Luna 5 µm Phenyl-Hexyl 100 Å, LC Column 250 x 4.6 mm, Ea

View Product

  • 1. Octahydro-1,3,5,7-tetranitro-1,3,5,7-tetrazocine (HMX)
  • 2. Hexahydro-1,3,5-trinitro-1,3,5-triazine
  • 3. 1,3,5-Trinitrobenzene
  • 4. Methyl-2,4,6-trinitrophenylnitramine (Tetryl)
  • 5. 1,3-Dinitrobenzene
  • 6. Nitrobenzene
  • 7. Nitroglycerin
  • 8. 2,4,6-Trinitrotoluene
  • 9. 3,5-Dinitroaniline
  • 10. 4-Amino-2,6 dinitrotoluene
  • 11. 2 Amino-4,6 dinitrotoluene
  • 12. 2,6-Dinitrotoluene
  • 13. 2,4-Dinitrotoluene
  • 14. Pentaerythritol tetranitrate
  • 15. 2-Nitrotoluene
  • 16. 4-Nitrotoluene
  • 17. 3-Nitrotoluene
Similar Applications