Ketones on Luna 5µm C18(2) 150 x 4.6 mm 0.714 mL/min

Ketones on Luna 5µm C18(2) 150 x 4.6 mm 0.714 mL/min
LC Conditions
Column

Luna 5 µm C18(2) 100 Å, LC Column 150 x 4.6 mm, Ea

Brand

Luna

Part No

00F-4252-E0

Phase Name

C18(2)

Elution Type

Gradient

Mobile Phase

A: water

B: acetonitrile

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

95

5

2

2.81

95

5

3

30.85

5

95

4

38

5

95

5

38.01

95

5

6

50.22

95

5

Flow Rate

0.71 mL/min

Temperature

45°C

Detection

UV-Vis Abs.-Diode Array (PDA) @ (25°C)

Detection Info
Sample Notes

SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Luna 5 µm C18(2) 100 Å, LC Column 150 x 4.6 mm, Ea
00F-4252-E0

Luna 5 µm C18(2) 100 Å, LC Column 150 x 4.6 mm, Ea

View Product

  • 1. Acetone
  • 2. 2-Butanone
  • 3. 2-Pentanone
  • 4. Acetophenone
  • 5. 2-Heptanone
  • 6. Butyrophenone
  • 7. 2-Nonanone
  • 8. Hexanophenone
  • 9. Octanophenone
  • 10. 2-Tridecanone
  • 11. Decanophenone
  • 12. 2-Pentadecanone
  • 13. 2-Hexadecanone
Similar Applications