Chlorinated Pollutants on ZB-5 by ECD
GC Conditions
Column
Zebron ZB-5, GC Cap. Column 60 m x 0.25 mm x 0.25 µm, Ea
Part No
7KG-G002-11
Phase Name
ZB-5
Injection
Pulsed @ 280°C
Carrier Gas
Constant Flow Helium, 2.6 mL/min
Detection
Electron Capture (ECD) @ (330°C)
Detection Info
Sample Notes
Analytes are 12 ppb Pulsed Splitless Injection @ 4 bar for 1 minute, splitless for 2 min. Application Courtesy of Jurgen Banger of Landeslabor Schleswig Holstein.
Order Items Used in This Application
- 1. 2-Brombiphenyl
- 2. Hexachlorobenzene
- 3. alpha-HCH (alpha-BHC)
- 4. y-HCH (g-BHC)
- 5. Moschus-Xylol
- 6. PCB 29
- 7. b-HCH (b-BHC)
- 8. PCB 28
- 9. d-HCH (d-BHC)
- 10. Aldrin
- 11. PCB 52
- 12. e-HCH
- 13. 4,4-Dichlorbenzophennone
- 14. cis-Heptachlorepoxide
- 15. PCB 101
- 16. cis-Chlordane
- 17. Dieldrin
- 18. PCB 118
- 19. PCB 153
- 20. p,p'-DDD
- 21. PCB 138
- 22. p,p'-DDT
- 23. PCB 167
- 24. PCB 180
- 25. PBDE47
- 26. PBDE100
- 27. PBDE99
- 28. PBDE116
- 29. PBDE85
- 30. PBDE154
- 31. PBDE153
Similar Applications