6-MAM from urine by Strata-XC on a ZB-Drug1
GC Conditions
Column
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Part No
7CD-G023-08
Phase Name
ZB-Drug-1
Injection
Split 10:1 @ 260°C
Carrier Gas
Constant Flow Helium, 0.3 mL/min
Detection
Mass Selective (MSD) @ (320°C)
Detection Info
Sample Notes
Analytes are 10ng/mL in urine derivatized with BSTFA Sim ions: (Quant ion*) D3-6-MAM (*402,403): 6-MAM (*399,400,340) Note: Method kept this way for continuity, the sample can be cleaned furthur by using different wash steps. For more information contact a Phenomenex technical consultant.
Sample Preparation Details
Condition 1
1 mL Methanol at 1 mL/min
Condition 2
1 mL Water at 1 mL/min
Condition 3
1 at 1
Load
Load 1.00 mL Sample diluted in Urine
Wash 1
1 mL 0.1 N HCl at 1 mL/min
Wash 2
1 mL Methanol at 1 mL/min
Wash 3
3 at 1
Dry
10minutes at > 10 inch Hg
Elution 1
1 mL Ammonium hydroxide in methanol 5% at 1 mL/min
Evaporation
Evaporate to dryness
Reconstitute
50 µL BSTFA/Pyridine