GC Conditions
Column
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Part No
7CD-G023-08
Phase Name
ZB-Drug-1
Injection
Split 10:1 @ 260°C
Carrier Gas
Constant Flow Helium, 0.7 mL/min
Detection
Mass Selective (MSD) @ (320°C)
Detection Info
Sample Notes
Analytes are 20ppm derivatized with HFBA Sim ions: (Quant ion*) D11-Amphetamine (*244,128): Amphetamine (*240,118,117): D14-Methamphetamine (*261,213): Methamphetamine (*254,210,118): D5-MDA (*167,380,244): MDA (*162,375,240): D5-MDMA (*258,213,164): MDMA (*254,210,389): D5-MDEA (*273,241,408): MDEA (*268,240,403): Phentermine (*254,214,132): Phenylpropanolamine (*240,330,275): Ephedrine (*254,210,344) Note: Amphetamine-d11, Phentermine and Phenylpropanolamine do not share quantitation or qualification ions allowing separation via the mass spectrometer.
Order Items Used in This Application
- 1. Phenylpropanolamine
- 2. Phentermine
- 3. Amphetamine-d11
- 4. Amphetamine
- 5. Ephedrine
- 6. Methamphetamine-d14
- 7. Methamphetamine
- 8. MDA-d5
- 9. MDA
- 10. MDMA-d5
- 11. MDMA
- 12. MDEA-d5
- 13. MDEA