EPA 531.1 Carbamate Pesticides at High Speed using Kinetex 2.6µm C18 50 x 2.1mm ID
LC Conditions
Official Method
EPA 531
Column
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Brand
Kinetex
Part No
00B-4462-AN
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: Water with 0.1% H3PO4
B: Acetonitrile with 0.1% H3PO4
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
95
5
2
3
5
95
Flow Rate
1.00 mL/min
Temperature
0°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Preparation Details
Description
Strata-X 33 µm Polymeric Reversed Phase, 30 mg / 3 mL, Tubes, 50/Pk
Part No
8B-S100-TBJ
Condition 1
1 mL Methanol
Condition 2
1 mL Loading buffer
Load
Load Sample
Wash 1
1 mL Loading buffer
Wash 2
1 mL Methanol/Water (15:85)
Dry
2 minutes at 10 mm Hg vacuum
Elution 1
0.5 mL 5% Formic acid in Methanol
Elution 2
0.5 mL 5% Formic acid in Methanol
Evaporation
Evaporate to dryness
Reconstitute
1 mL Mobile phase
Order Items Used in This Application
- 1. Aldicarb-sulfoxide
- 2. Aldicarb-sulfone
- 3. Oxamyl
- 4. Methomyl
- 5. 3-Hydroxy-Carbofuran
- 6. Aldicarb sulfone-related impurity
- 7. Aldicarb
- 8. Baygon (Propoxur)
- 9. Carbofuran
- 10. Carbaryl
- 11. 1-Napthol
- 12. Methiocarb
Similar Applications