6-MAM, PCP and Methadone on Kinetex 2.6µm C18 50 x 2.1mm ID with Acquity UPLC MS/MS (1 mL/min)

6-MAM, PCP and Methadone on Kinetex 2.6µm C18 50 x 2.1mm ID with Acquity UPLC MS/MS (1 mL/min)
LC Conditions
Column

Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea

Brand

Kinetex

Part No

00B-4462-AN

Phase Name

C18

Elution Type

Gradient

Mobile Phase

A: 0.1% Formic acid in H2O

B: 0.1% Formic acid in ACN

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

90

10

2

1

5

95

3

1.4

5

95

4

1.41

5

95

5

2

90

10

Flow Rate

1.00 mL/min

Temperature

0°C

Detection

Mass Spectrometer (MS) @ (110°C)

Detection Info

Order Items Used in This Application

Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
00B-4462-AN

Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea

View Product

  • 1. 6-MAM
  • 2. 6-MAM-d3
  • 3. PCP
  • 4. PCP-d5
  • 5. Methadone
  • 6. Methadone-d9