PCP on Kinetex 2.6µm C18 50 x 2.1mm ID by LCMS

PCP on Kinetex 2.6µm C18 50 x 2.1mm ID by LCMS
LC Conditions
Column

Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea

Brand

Kinetex

Part No

00B-4462-AN

Phase Name

C18

Elution Type

Gradient

Mobile Phase

A: 0.1% Formic acid in H2O

B: 0.1% Formic acid in ACN

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

97

3

2

0.25

97

3

3

2.75

5

95

4

3

5

95

5

3.01

97

3

6

4.5

97

3

Flow Rate

500.00 mL/min

Temperature

0°C

Detection

Mass Spectrometer (MS) @ (22°C)

Detection Info
Sample Preparation Details
Description

Strata-X-C 33 µm Polymeric Strong Cation, 30 mg / 3 mL, Tubes, 50/Pk

Part No

8B-S029-TBJ

Condition 1

1 mL methanol at 1 mL/min

Condition 2

1 mL water at 1 mL/min

Load

Load 1.00 mL Sample diluted in 1

Wash 1

1 mL 0.1N hydrochloric acid at 2 mL/min

Wash 2

1 mL Methanol at 2 mL/min

Elution 1

0.5 mL 5% NH4OH/methanol at 1 mL/min

Elution 2

0.5 mL 5% NH4OH/methanol at 1 mL/min

Evaporation

Evaporate to dryness

Reconstitute

200 µL Initial LC mobile phase (97/3)

Order Items Used in This Application

Strata-X-C 33 µm Polymeric Strong Cation, 30 mg / 3 mL, Tubes, 50/Pk
8B-S029-TBJ

Strata-X-C 33 µm Polymeric Strong Cation, 30 mg / 3 mL, Tubes, 50/Pk

View Product

Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
00B-4462-AN

Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea

View Product

  • 1. PCP
  • 2. PCP-d5
Similar Applications

No Similar applications found.