Extraction of 6-MAM using Strata X-C on Kinetex 2.6µm C18 50 x 2.1mm ID by LCMS

Extraction of 6-MAM using Strata X-C on Kinetex 2.6µm C18 50 x 2.1mm ID by LCMS
LC Conditions
Column

Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea

Brand

Kinetex

Part No

00B-4462-AN

Phase Name

C18

Elution Type

Gradient

Mobile Phase

A: 0.1% Formic acid in H2O

B: 0.1% Formic acid in ACN

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

97

3

2

0.24

97

3

3

3.02

5

95

4

3.3

5

95

5

3.31

97

3

6

4.97

97

3

Flow Rate

450.00 mL/min

Temperature

0°C

Detection

Mass Spectrometer (MS) @ (22°C)

Detection Info
Sample Preparation Details
Description

Strata-X-C 33 µm Polymeric Strong Cation, 30 mg / 3 mL, Tubes, 50/Pk

Part No

8B-S029-TBJ

Condition 1

1 mL Methanol at 2 mL/min

Condition 2

1 mL Water at 2 mL/min

Load

Load 1.00 mL Sample diluted in 1

Wash 1

1 mL 0.1N HCl at 2 mL/min

Wash 2

1 mL Methanol at 2 mL/min

Dry

2-3 mins at 10 of Hg

Elution 1

0.5 mL 5% ammonium hydroxide/methanol at 1 mL/min

Elution 2

0.5 mL 5% ammonium hydroxide/methanol at 1 mL/min

Evaporation

Evaporate to dryness

Reconstitute

200 µL Initial LC mobile phase (A/B=97/3)

Order Items Used in This Application

Strata-X-C 33 µm Polymeric Strong Cation, 30 mg / 3 mL, Tubes, 50/Pk
8B-S029-TBJ

Strata-X-C 33 µm Polymeric Strong Cation, 30 mg / 3 mL, Tubes, 50/Pk

View Product

Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
00B-4462-AN

Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea

View Product

  • 1. 6 MAM
  • 2. 6-MAM-d3
Similar Applications

No Similar applications found.