Extraction of 6-MAM using Strata X-C on Kinetex 2.6µm C18 50 x 2.1mm ID by LCMS
LC Conditions
Column
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Brand
Kinetex
Part No
00B-4462-AN
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 0.1% Formic acid in H2O
B: 0.1% Formic acid in ACN
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
97
3
2
0.24
97
3
3
3.02
5
95
4
3.3
5
95
5
3.31
97
3
6
4.97
97
3
Flow Rate
450.00 mL/min
Temperature
0°C
Detection
Mass Spectrometer (MS) @ (22°C)
Detection Info
Sample Preparation Details
Description
Strata-X-C 33 µm Polymeric Strong Cation, 30 mg / 3 mL, Tubes, 50/Pk
Part No
8B-S029-TBJ
Condition 1
1 mL Methanol at 2 mL/min
Condition 2
1 mL Water at 2 mL/min
Load
Load 1.00 mL Sample diluted in 1
Wash 1
1 mL 0.1N HCl at 2 mL/min
Wash 2
1 mL Methanol at 2 mL/min
Dry
2-3 mins at 10 of Hg
Elution 1
0.5 mL 5% ammonium hydroxide/methanol at 1 mL/min
Elution 2
0.5 mL 5% ammonium hydroxide/methanol at 1 mL/min
Evaporation
Evaporate to dryness
Reconstitute
200 µL Initial LC mobile phase (A/B=97/3)
Order Items Used in This Application
- 1. 6 MAM
- 2. 6-MAM-d3
Similar Applications