Atenolol and Related Substances CRS using Luna 5um C18(2) 125 x 4.0mm
LC Conditions
Column
Luna 5 µm C18(2) 100 Å, LC Column 125 x 4 mm, Ea
Brand
Luna
Part No
00E-4252-D0
Phase Name
C18(2)
Elution Type
Isocratic
Mobile Phase
A: EP Mobile Phase
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
100
0
Flow Rate
0.60 mL/min
Temperature
25°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
12.5 mM KH2PO4 in Water pH 3.0 + 2.0g Sodium Octanesulfonate + 0.8g Tetrabutyl Ammonium Hydrogen Sulfate/Methanol/THF v/v% 80:18:2
Order Items Used in This Application
- 1. Impurity B
- 2. Impurity A
- 3. Impurity J
- 4. Impurity I
- 5. impurities D and E
- 6. Impurity F
- 7. Impurity G
- 8. Impurity H