Extraction of Expanded Amphetamines Using Strata X-C on Kinetex 2.6µm C18 50 x 2.1mm ID by LCMS
LC Conditions
Column
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Brand
Kinetex
Part No
00B-4462-AN
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 0.1%FA-5mM ammonium acetate in H2O
B: 0.1%FA-5mM ammonium acetate in Methanol
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
90
10
2
3
60
40
3
3.02
5
95
4
3.3
5
95
5
3.31
90
10
6
5.1
90
10
Flow Rate
450.00 mL/min
Temperature
0°C
Detection
Mass Spectrometer (MS) @ (22°C)
Detection Info
Sample Preparation Details
Description
Strata-X-C 33 µm Polymeric Strong Cation, 30 mg / 3 mL, Tubes, 50/Pk
Part No
8B-S029-TBJ
Pretreatment Note
Blown down samples (after evaporation) were reconstituted in 1 mL of initial mobile phase (A/B=90/10).
Condition 1
1 mL Methanol at 2 mL/min
Condition 2
1 mL Water at 2 mL/min
Load
Load 1.00 mL Sample diluted in 100 mM KH2PO4 (pH 6)
Wash 1
1 mL 0.1 N HCL at 2 mL/min
Wash 2
1 mL Methanol at 2 mL/min
Dry
2 minutes at 10 of Hg pressure
Elution 1
0.5 mL 5% Ammonium hydroxide/methanol at 1 mL/min
Elution 2
0.5 mL 5% Ammonium hydroxide/methanol at 1 mL/min
Evaporation
Evaporate to dryness
Reconstitute
1000 µL Initial mobile phase (A/B=90/10)
Order Items Used in This Application
- 1. HMMA
- 2. Amphetamine-d11
- 3. Amphetamine
- 4. Methamphetamine
- 5. Methamphetamine-d14
- 6. MDA-d5
- 7. MDA
- 8. MDMA-d5
- 9. MDMA
- 10. MDEA-d5
- 11. MDEA
Similar Applications