Extraction of Benzodiazepines Using Strata Screen-C on Kinetex 2.6µm C18 50 x 2.1mm ID by LCMS
LC Conditions
Column
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Brand
Kinetex
Part No
00B-4462-AN
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 0.1%FA in H2O
B: 0.1%FA in Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
65
35
2
2
65
35
3
2.01
5
95
4
2.3
5
95
5
2.31
65
35
6
4.1
65
35
Flow Rate
450.00 mL/min
Temperature
0°C
Detection
Mass Spectrometer (MS) @ (22°C)
Detection Info
Sample Preparation Details
Description
Strata Screen-C (55 µm, 70 Å), 100 mg / 3 mL, Tubes, 50/Pk
Part No
8B-S016-EBJ
Condition 1
3 mL Methanol at 1 mL/min
Condition 2
3 mL Water at 1 mL/min
Condition 3
1 mL 100 mM phosphate buffer, pH 6 (adjust with phosphoric acid) at 1 mL/min
Load
Load 2.00 mL Sample diluted in 100 mm acetate buffer
Wash 1
2 mL Water at 1 mL/min
Wash 2
2 mL 20% acetonitrile/100 mm phosphate buffer, pH 6 at 1 mL/min
Dry
2-3 mins at 10 of Hg
Elution 1
2 mL Ethylacetate at 1 mL/min
Elution 2
2 mL Ethylacetate at 1 mL/min
Evaporation
Evaporate to dryness
Reconstitute
1 mL Initial mob phase(A/B=65/35)
Order Items Used in This Application
- 1. alpha-Hydroxy alprazolam
- 2. alpha-HydroxyAlprazolam-d5
- 3. Oxazepam-d5
- 4. Oxazepam
- 5. Nordiazepam-d5
- 6. Nordiazepam
- 7. Lorazepam
- 8. Clonazepam
- 9. Temazepam-d5
- 10. Temazepam
- 11. Diazepam
Similar Applications