Clarity OTX- Recovery Data- MS purity data of 27mer P-S peak from 18868
LC Conditions
Column
Clarity 3 µm Oligo-RP, LC Column 150 x 2 mm, Ea
Brand
Clarity LC
Part No
00F-4441-B0
Phase Name
Oligo-RP
Elution Type
Gradient
Mobile Phase
A: 200mM HFIP/ 14mM TEA/ 2% MeOH/ water
B: 200mM HFIP/ 14mM TEA/ 98% MeOH/ water
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
90
10
2
5
85
15
3
20
35
65
4
20.1
5
95
Flow Rate
0.25 mL/min
Temperature
40°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Clarity Oligo-RP 4 x 2.0mm ID, 10/Pk Part No.: AJ0-8134 - Holder Part No.: KJ0-4282
Sample Preparation Details
Description
Clarity OTX 100 mg Plate, 100 mg/well 96-well plate, Ea
Part No
8E-S103-EGA
Condition 1
0.5 mL MeOH at 1 mL/min
Condition 2
0.5 mL Clarity OTX Equlibration Buffer at 1 mL/min
Sample Pretreatment
O
Load
Load 500.00 µL Sample diluted in 1:1 Clarity OTX loading buffer/ sample
Wash 1
0.5 mL Clarity OTX wash solvent at 1 mL/min
Wash 2
0.5 mL Clarity OTX equilibration Buffer at 1 mL/min
Dry
N2 at 10u vaccuum
Elution 1
0.5 mL Clarity OTX elution buffer at 1 mL/min
Evaporation
Evaporate to dryness
Reconstitute
100 µL Aobile phase A
Order Items Used in This Application
- 1. 27 mer P-S Oligonucleotide