Clarity OTX- MS Sensitivity of recovered 21mer at 500ng/mL from plasma
LC Conditions
Column
Clarity 3 µm Oligo-RP, LC Column 50 x 2 mm, Ea
Brand
Clarity LC
Part No
00B-4441-B0
Phase Name
Oligo-RP
Elution Type
Gradient
Mobile Phase
A: 1%HFIPA/ 0.1%DIEA/ water
B: 90% MeOH/ 0.075%HFIPA/ 0.0375% DIEA
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
90
10
2
20
65
35
Flow Rate
0.25 mL/min
Temperature
60°C
Detection
Mass Spectrometer (MS) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Clarity Oligo-RP 4 x 2.0mm ID, 10/Pk Part No.: AJ0-8134 - Holder Part No.: KJ0-4282
Sample Preparation Details
Description
Clarity OTX 100 mg Plate, 100 mg/well 96-well plate, Ea
Part No
8E-S103-EGA
Condition 1
0.5 mL MeOH at 1 mL/min
Condition 2
0.5 mL Clarity OTX Equilibration Buffer at 1 mL/min
Load
Load 500.00 µL Sample diluted in OTX loading buffer/sample
Wash 1
0.5 mL Clarity OTX Wash Buffer at 1 mL/min
Wash 2
0.5 mL Clarity OTX Equilibration Buffer at 1 mL/min
Dry
10 um vaccuum
Elution 1
0.5 mL Clarity OTX Elution Buffer at 1 mL/min
Evaporation
Evaporate to dryness
Reconstitute
200 µL Mobile phase A
Order Items Used in This Application
- 1. 21mer DNA oligo