Clarity OTX- Sensitivity of 19mer P-S Oligo from Liver Tissue

Clarity OTX- Sensitivity of 19mer P-S Oligo from Liver Tissue
LC Conditions
Column

Clarity 3 µm Oligo-RP, LC Column 50 x 2 mm, Ea

Brand

Clarity LC

Part No

00B-4441-B0

Phase Name

Oligo-RP

Elution Type

Gradient

Mobile Phase

A: 200mM HFIP/ 14mM TEA/ 2% MeOH/ water

B: 200mM HFIP/ 14mM TEA/ 98% MeOH/ water

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

90

10

2

10

35

65

Flow Rate

0.25 mL/min

Temperature

40°C

Detection

UV-Vis Abs.-Variable Wave.(UV) @ (22°C)

Detection Info
Sample Notes

SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Clarity Oligo-RP 4 x 2.0mm ID, 10/Pk Part No.: AJ0-8134 - Holder Part No.: KJ0-4282

Sample Preparation Details
Description

Clarity OTX 100 mg Plate, 100 mg/well 96-well plate, Ea

Part No

8E-S103-EGA

Condition 1

0.5 mL MeOH at 1 mL/min

Condition 2

0.5 mL Clarity OTX Equlibration Buffer at 1 mL/min

Sample Pretreatment

O

Load

Load 500.00 µL Sample diluted in 1:1 Clarity OTX loading buffer/ sample

Wash 1

0.5 mL Clarity OTX wash solvent at 1 mL/min

Wash 2

0.5 mL Clarity OTX equilibration Buffer at 1 mL/min

Dry

N2 at 10u vaccuum

Elution 1

0.5 mL Clarity OTX elution buffer at 1 mL/min

Evaporation

Evaporate to dryness

Reconstitute

100 µL Aobile phase A

Order Items Used in This Application

Clarity 3 µm Oligo-RP, LC Column 50 x 2 mm, Ea
00B-4441-B0

Clarity 3 µm Oligo-RP, LC Column 50 x 2 mm, Ea

View Product

Clarity OTX, 100 mg / well, 96-Well Plates, Ea
8E-S103-EGA

Clarity OTX, 100 mg / well, 96-Well Plates, Ea

View Product

  • 1. 19 mer P-S oligonucleotide