Paroxetine Hydrochloride Analysis (HP. Eur) on Luna C18(2) 5µ 250 x 4.6mm ID

Paroxetine Hydrochloride Analysis (HP. Eur) on Luna C18(2) 5µ 250 x 4.6mm ID
LC Conditions
Column

Luna 5 µm C18(2) 100 Å, LC Column 250 x 4.6 mm, Ea

Brand

Luna

Part No

00G-4252-E0

Phase Name

C18(2)

Elution Type

Gradient

Mobile Phase

A: 5:100:900 TFA/THF/H2O

B: 5:100:900 TFA/THF/CAN

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

80

20

2

30

80

20

3

50

20

80

4

55

20

80

5

60

80

20

6

65

80

20

Flow Rate

1.50 mL/min

Temperature

40°C

Detection

UV-Vis Abs.-Variable Wave.(UV) @ (22°C)

Detection Info

Order Items Used in This Application

Luna 5 µm C18(2) 100 Å, LC Column 250 x 4.6 mm, Ea
00G-4252-E0

Luna 5 µm C18(2) 100 Å, LC Column 250 x 4.6 mm, Ea

View Product

  • 1. Impurity B
  • 2. Impurity A
  • 3. Impurity E
  • 4. Paroxetine hydrochloride
  • 5. Impurity C
Similar Applications