Paroxetine Hydrochloride Analysis (HP. Eur) on Luna C18(2) 5µ 250 x 4.6mm ID
LC Conditions
Column
Luna 5 µm C18(2) 100 Å, LC Column 250 x 4.6 mm, Ea
Brand
Luna
Part No
00G-4252-E0
Phase Name
C18(2)
Elution Type
Gradient
Mobile Phase
A: 5:100:900 TFA/THF/H2O
B: 5:100:900 TFA/THF/CAN
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
80
20
2
30
80
20
3
50
20
80
4
55
20
80
5
60
80
20
6
65
80
20
Flow Rate
1.50 mL/min
Temperature
40°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Order Items Used in This Application
- 1. Impurity B
- 2. Impurity A
- 3. Impurity E
- 4. Paroxetine hydrochloride
- 5. Impurity C
Similar Applications