Expanded Amphetamines on Kinetex C18 2.6µm 50x2.1mm by LCMS

Expanded Amphetamines on Kinetex C18 2.6µm 50x2.1mm by LCMS
LC Conditions
Column

Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea

Brand

Kinetex

Part No

00B-4462-AN

Phase Name

C18

Elution Type

Gradient

Mobile Phase

A: 0.1%FA-5mM ammonium acetate in H2O

B: 0.1%FA-5mM ammonium acetate in Methanol

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

90

10

2

3

60

40

3

3.02

5

95

4

3.3

5

95

5

3.31

90

10

6

5.1

90

10

Flow Rate

450.00 mL/min

Temperature

0°C

Detection

Mass Spectrometer (MS) @ (22°C)

Detection Info

Order Items Used in This Application

Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
00B-4462-AN

Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea

View Product

  • 1. HMMA
  • 2. Amphetamine-d11
  • 3. Amphetamine
  • 4. Methamphetamine
  • 5. Methamphetamine-d14
  • 6. MDA-d5
  • 7. MDA
  • 8. MDMA-d5
  • 9. MDMA
  • 10. MDEA-d5
  • 11. MDEA
Similar Applications