Pain Panel on Kinetex C18 2.6µm 50 x 2.1 mm by LC/MS
LC Conditions
Column
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Brand
Kinetex
Part No
00B-4462-AN
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 0.1% Formic acid in H2O
B: 0.1% Formic acid ACN
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
95
5
2
0.5
95
5
3
2.5
5
95
4
3.5
5
95
5
3.51
95
5
6
5.3
95
5
Flow Rate
450.00 mL/min
Temperature
0°C
Detection
Mass Spectrometer (MS) @ (22°C)
Detection Info
Order Items Used in This Application
- 1. Morphine
- 2. Oxymorphone
- 3. Hydromorphone
- 4. Codeine
- 5. Naloxone
- 6. Oxycodone
- 7. Naltrexone
- 8. Hydrocodone
- 9. Tramadol
- 10. Meperidine
- 11. Fentanyl
- 12. Buprenorphine
- 13. Methadone
Similar Applications