USP Glycerin Purity Assay on ZB-BAC-2
GC Conditions
Column
Zebron ZB-BAC-2, GC Cap. Column 30 m x 0.53 mm x 2.00 µm, Ea
Part No
7HK-G022-32
Phase Name
ZB-BAC-2
Injection
Split 10:1 @ 220°C
Carrier Gas
Constant Flow Helium, 4.5 mL/min
Detection
Flame Ionization (FID) @ (250°C)
Detection Info
Sample Notes
Concentrations were 2.0 mg/mL Glycerin, 0.050 mg/mL ethylene glycol, 0.050 mg/mL diethylene glycol, and 0.10 mg/mL 2,2,2-trichloroethanol (IS) Run as a dual column analysis with ZB-BAC-1 Resolution = 11.09 between glycerin and diethylene glycol
Order Items Used in This Application
- 1. Ethylene glycol
- 2. 2,2,2-Trichloroethanol
- 3. Diethylene glycol
- 4. Glycerin
Similar Applications
No Similar applications found.